Installing the Perforated Shelves

[1]
Rear Pull-out Stop
[2]
Tilt Protection Device
[3]
Front Pull-out Stop
[4]
Shelf

Figure 5-9 Perforated Shelf

1. Push the shelf [4] onto the shelf support brackets with the tilt protection devices [2] facing
the rear panel of the incubator.
2. Slightly raise the perforated shelf so that the pull-out stops [1] and [3] can slide over the
shelf support brackets.
3. Make sure that the shelves and both of their tilt protection devices are free to move over
the shelf support brackets.

Mains connection

The incubator has a class I, protection-earthed enclosure. To minimize the risk of electrical
shock, use the AC power cord supplied to connect the incubator to a correctly installed and
protection-earthed power supply source, with the following features in place for each
incubator:
• T 16 A slow-blow fusing
• B 16 circuit breaker
Thermo Scientific
Contact with live electrical components may cause a lethal electric
shock.
Before connecting the incubator to the power supply, check the
power cord and the plug for damage. Do not use damaged cables
for connecting the incubator to the power supply!
Electric shock
Installation procedures
Mains connection