Space Requirement

For built-in units following clearances should be kept:
A, B, C and D see Operating Instructions of the unit.
E (mm/inch)
100 / 4

Installation Built-in Units

IGS
IMH/IMH-S
OGS/OMS
OGH/OGH-S
OMH/OMH-S
When installing the incubator, make sure that the installation and supply connections remain
freely accessible.
The specified side clearances represent minimum distances.
Thermo Scientific
F (mm/inch)
G (mm/inch)
50 / 2
80 / 3.2
Clearances without exhaust hose.
Clearances, close air port with plug, supplied with the unit.
Clearances, with additional space to operate to air slide, with air exhaust system
and exhaust hose Ø 40 mm (1.58 in)/1.5 m (59 in) (original Thermo accessory),
shortened to required length.
Clearances, with air exhaust system and exhaust hose Ø 40 mm (1.58 in)/1.5 m
(59 in) (original Thermo accessory), shortened to required length. Operating
temperature up to max. 250 °C (482 °F).
After connecting the built-in unit to the electrical
power supply, avoid damaging the power cord, then
sliding into the installation position.
Fig. 3
H (mm/inch)
30 / 1.2
Installation
Room Ventilation
Heratherm
3-3