Maintenance

Service Intervals

Service Intervals
During running operation, the following service works must be performed:

3-monthly service

• Perform the comparative temperature measurement outlined in the following section.

Annual Service

• Have the incubator inspected and services by a Thermo Scientific authorized Technical
Service agent.
10-2
Heratherm
If safety devices were removed or disabled for inspections, the
incubator must not be operated before the safety devices have been
reinstalled and checked for their correct function.
To avoid major malfunctions of the incubator and associated
safety hazards that may result in death, serious injuries, or
damage to the incubator and other equipment, use spare parts
approved by Thermo Scientific only. Third-party spares without
Thermo Scientific approval void the limited warranty.
Do not modify the incubator in any way without obtaining the
prior written authorization from Thermo Scientific. Unauthorized
modifications may compromise operational safety and give rise
to hazards that may result in death, serious injuries, or damage to
the incubator and other equipment.
Thermo Electron LED GmbH offer a product-specific service contract
comprising all necessary tests and service work.
Functional Check
Spare Parts and User Modifications
Service contract