Maintenance
Maintenance and inspection at regular intervals of the features and components listed below
are mission-critical to maintain the product in a fully operative and safe condition and avoid
malfunctions due to ageing and wear. Failure to perform maintenance on a regular basis may
result in:
• deviations in heating performance
• damage to samples
• loss of control over temperature distribution throughout the work space

Inspections and Checks

To ensure the operational performance and safety of the incubator and its functions, the
components listed below must be checked at regular intervals.

Regular Checks

• Check the incubator for overall cleanliness and remove any residues from previous
processes.
• Check the locking screw for the glass panel
• To avoid incubator operation without an appropriate fresh air supply, check the air filter
(optional) in the air inlet path for contamination.

Semi-annual Inspection

• Check integrity and proper seating of the seal.
• Swap the fresh-air filter (accessory) in air inlet.
• Perform functional check of the control panel and of the incubator's built-in controller.
• Perform electrical safety check in accordance with the relevant national regulations