Connecting The Processor; Making The Electrical Connections; Connecting The Jumpers At Tb2 And Tb3 And The Wires To Tb1 - Kodak X-Omat 5000 RA PROCESSOR Installation Instructions Manual

Hide thumbs Also See for X-Omat 5000 RA PROCESSOR:
Table of Contents

Advertisement

INSTALLATION INSTRUCTIONS
Section 3: Connecting the PROCESSOR

Making the Electrical Connections

Connecting the JUMPERS at TB2 and TB3 and the Wires to TB1

ELECTRICAL
BOX
H148_0183ACA
H148_0183AA
20
[1] Extend the ELECTRICAL BOX.
COMPONENT
[2] Remove the COMPONENT PLATE COVER.
PLATE
COVER
[3] Determine the supply voltage to be applied to the
[4] Install the correct ELECTRIC SERVICE LABEL
H148_0185CCA
H148_0185CA
PROCESSOR.
from the PREPACK onto the DATA PLATE.
DATA
PLATE
KODAK
X-OMAT 5000 RA Processor
SERVICE
SERIAL
3425
CODE
NUMBER
115B
LISTED
Made in USA by
Eastman Kodak Company
Rochester, New York 14650
apply ELECTRIC
SERVICE LABEL here
January 1996 – 5B6339