7.1.1.3
Specifying the Sensor Type for CERAVAC Sensors
When connecting all CERAVAC sensors
through the connections C1, D1 and E1 on the rear of the instrument
26) then during first-time commissioning the status message Range? and as sensor type
CTR? is displayed in the left bottom area of the display field for the measurement channel.
Also when connecting the following CERAVAC sensors through connections C2, D2 and E2
on the rear of the instrument
the status message Range? and as sensor type CTR? is displayed in the left bottom area
of the display field for the measurement channel:
CTR90-0.1Torr
CTR90-1Torr
CTR90-10Torr
CTR90-20Torr
CTR90-100Torr
CTR90-1000Torr
To specify the connected sensor proceed as follows:
Press in the measurement mode the touchscreen surface for approximately 1 second.
o The main menu with an overview of the parameter groups is displayed.
Tap in the main menu on the parameter group of the desired channel.
o The parameters available for this parameter group are displayed.
GA300550402_002_C1 – 10/2016
(
(
Figure 7, page 26), then during first-time commissioning
CTR91-0.1Torr
CTR91-1Torr
CTR91-10Torr
CTR91-20Torr
CTR91-100Torr
CTR91-1000Torr
Figure 62 – Display of sensor type CTR?
Figure 63 – Opening the main menu
Figure 64 – Selection of parameter group Channel
Chapter 3.2 Suitable Sensors, page 15)
(
Figure 7, page