HP 8719/20/22ET Specifications and Characteristics
Instrument Specifications
Table 2-6 2.4-mm (50 Ω) Device Connector Type
HP 8722ET (2.4-mm, 50 Ω)
Description
50 to 500 MHz
a
Directivity
(dB)
a
Source Match
(dB)
a
Load Match
(dB)
b
Reflection
Tracking (dB)
a
Transmission
Tracking (dB)
Tracking
Stability
(Ratio
Measurement)
Magnitude
0.02/˚C, typ.
(dB)
Phase (deg)
0.1/˚C, typ.
c
Crosstalk
(dB)
a. Does not include effects of cable on test ports.
b. Does not include rolloff below 500 MHz which is typically −18 dB at 100 MHz, and −25 dB
at 50 MHz.
c. Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.
2-10
0.5 to 2 GHz
23
23
16
20
22
20
± 3
± 3
± 3
± 3
0.03/˚C, typ.
0.1/˚C, typ.
95
95
Specification
2 to 8 GHz
8 to 20 GHz
21
14
20
± 3
± 3
0.03/˚C, typ.
0.04/˚C, typ.
0.3/˚C, typ.
0.5/˚C, typ.
95
20 to 40 GHz
16
15
11
10
20
15
± 3
± 4
± 3
± 3
0.06/˚C, typ.
1.0/˚C, typ.
88
82