Table 1-5 3.5-mm Device Connector Type
Network Analyzer: HP 8719ES/20ES, Option 400
Calibration Kit: HP 85052C (3.5-mm, 50 Ω)
Cables: HP 85131F
Calibration: TRL
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .02/°C)
Magnitude (dB)
±(0.040 + 0.1/°C)
Phase (deg)
Trans. Tracking
±(0.013 + .02/°C)
Magnitude (dB)
±(0.086 + 0.1/°C)
Phase (deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
50 to 500 MHz
0.5 to 2 GHz
48
40
48
±(0.006 + .03/°C)
±(0.040 + 0.1/°C)
±(0.017 + .03/°C)
±(0.112 + 0.1/°C)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Specification
2 to 8 GHz
48
50
40
50
48
50
±(0.005 + .03/°C)
±(0.033 + 0.3/°C)
±(0.016 + .03/°C)
±(0.107 + 0.3/°C)
8 to 20 GHz
50
50
50
±(0.005 + .04/°C)
±(0.033 + 0.5/°C)
±(0.019 + .04/°C)
±(0.125 + 0.5/°C)