Table 1-11 Type-N Device Connector Type
Network Analyzer: HP 8719ES/20ES, Standard
Calibration Kit: HP 85054B (Type-N, 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.005 + .02/°C)
Magnitude (dB)
±(0.033 + 0.1/°C)
Phase (deg)
Trans. Tracking
±(0.014 + .02/°C)
Magnitude (dB)
±(0.092 + 0.1/°C)
Phase(deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
50 to 500 MHz
0.5 to 2 GHz
48
45
48
±(0.005 + .03/°C)
±(0.033 + 0.1/°C)
±(0.015 + .03/°C)
±(0.099 + 0.1/°C)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Specification
2 to 8 GHz
48
42
45
36
48
42
±(0.006 + .03/°C)
±(0.040 + .30/°C)
±(0.055 + .03/°C)
±(0.363 + .30/°C)
8 to 18 GHz
42
32
42
±(0.015 + .04/°C)
±(0.99 + 0.50/°C)
±(0.093 + .04/°C)
±(0.614 + .50/°C)