DELAY/THRU
DELETE
DELETE ALL FILES
DELETE FILE
DELTA LIMITS
DEMOD: AMPLITUDE
DEMOD: OFF
DEMOD: PHASE
DIRECTORY SIZE
DISK UNIT NUMBER
DISP MKRS ON off
defines the standard type as a transmission line of
specified length, for calibrating transmission
measurements.
deletes the segment indicated by the pointer.
deletes all files.
deletes a selected file.
sets the limits an equal amount above and below a
specified middle value, instead of setting upper and lower
limits separately. This is used in conjunction with
or
MIDDLE VALUE
for testing a device that is specified at a particular value
plus or minus an equal tolerance.
For example, a device may be specified at 0 dB ±3 dB.
Enter the delta limits as 3 dB and the middle value as
0 dB.
(Option 010 only) amplitude demodulation for CW time
transform measurements.
(Option 010 only) turns time domain demodulation off.
(Option 010 only) phase demodulation for CW TIME
transform measurements.
lets you specify the number of directory files to be
initialized on a disk. This is particularly useful with a
hard disk, where you may want a directory larger than the
default 256 files, or with a floppy disk you may want to
reduce the directory to allow extra space for data files. The
number of directory files must be a multiple of 8. The
minimum number is 8, and there is no practical maximum
limit. Set the directory size before initializing a disk.
specifies the number of the disk unit in the disk drive that
is to be accessed in an external disk store or load routine.
This is used in conjunction with the HP-IB address of the
disk drive, and the volume number, to gain access to a
specific area on a disk. The access hierarchy is HP-IB
address, disk unit number, disk volume number.
displays response and stimulus values for all markers that
are turned on. Available only if no marker functions are
on, for example
MKR STATS
Hardkey/Softkey Reference
MARKER
MIDDLE
.
Analyzer Functions
, to set limits
5-17