HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Table 1-8 7-mm Device Connector Type
Network Analyzer: HP 8719ES/20ES, Standard
Calibration Kit: HP 85050B (7-mm, 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .02/°C)
Magnitude (dB)
±(0.040 + 0.1/°C)
Phase (deg)
Trans. Tracking
Magnitude (dB)
±(0.066 + 0.1/°C)
Phase (deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
1-10
50 to 500 MHz
0.5 to 2 GHz
52
48
51
±(0.006 + .03/°C)
±(0.040 + 0.1/°C)
±(0.01 + .02/°C)
±(0.011 + .03/°C)
±(0.073 + 0.1/°C)
Specification
2 to 8 GHz
52
52
48
44
51
51
±(0.017 + .03/°C)
±(0.112 + 0.3/°C)
±(0.022 + .03/°C)
±(0.145 + 0.3/°C)
8 to 18 GHz
52
41
51
±(0.047 + .04/°C)
±(0.310 + 0.5/°C)
±(0.034 + .04/°C)
±(0.224 + 0.5/°C)