If correction is on at the time of an external store, the calibration set is stored to disk.
(Note that inactive calibrations are not stored to disk.) When an instrument state is loaded
into the analyzer from disk, the stimulus and response parameters are restored first. If
correction is on for the loaded state, the analyzer will load a calibration set from disk that
carries the same title as the one stored for the instrument state.

Conserving Memory

If you are concerned about conserving memory, either internal memory or external disk
space, some of the most memory-intensive operations include:
• two-port error correction
• interpolated error correction
• 1601 measurement points
• using time domain
• saving data arrays and graphics with the instrument state

Using Saved Calibration Sets

When you are saving to internal memory (CMOS, non-volatile memory), calibration sets
are linked to the instrument state and measurement parameter for which the calibration
was done. Therefore a saved calibration can be used for multiple instrument states as long
as the measurement parameter, frequency range, and number of points are the same. A
full 2-port calibration is valid for any S-parameter measurement with the same frequency
range and number of points. When an instrument state is deleted from memory, the
associated calibration set is also deleted if it is unused by any other state.
The following hints will help you avoid potential problems:
• If a measurement is saved with calibration and interpolated calibration on, it will be
restored with interpolated calibration on.
• A calibration stored from one instrument and recalled by a different one will be invalid.
To ensure maximum accuracy, always recalibrate in these circumstances.
• No record is kept in memory of the temperature when a calibration set was stored.
Instrument characteristics change as a function of temperature, and a calibration
stored at one temperature may be inaccurate if recalled and used at a different
temperature. Refer to
Characteristics,"
and
for allowable temperature ranges for individual specifications.
Chapter 1 , "HP 8719/20/22ES Specifications and
Chapter 2 , "HP 8719/20/22ET Specifications and Characteristics,"
Preset State and Memory Allocation
Memory Allocation
8-19

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HP 8719ET and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Hp 8720etHp 8719esHp 8722etHp 8720es8722es

Table of Contents