HP 8719ET Reference Manual page 27

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

Table 1-14 3.5-mm Device Connector Type
Network Analyzer: HP 8722ES, Standard
Calibration Kit: HP 85052D (3.5-mm, 50 Ω)
Cables: HP 85131F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
±(0.006 + .03/°C)
Magnitude (dB)
±(0.040 + 0.1/°C)
Phase (deg)
Trans. Tracking
±(0.026 + .03/°C)
Magnitude (dB)
±(0.172 + 0.1/°C)
Phase(deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
0.05 to 2 GHz
2 to 8 GHz
42
37
42
±(0.006 + .03/°C)
±(0.040 + 0.30/°C)
±(0.071 + .03/°C)
±(0.469 + 0.3/°C)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8722ES)
Specification
8 to 20 GHz
38
36
31
28
38
36
±(0.009 + .04/°C)
±(0.059 + 0.5/°C)
±(0.121 + .04/°C)
±(0.792 + 0.5/°C)
20 to 26.5 GHz
30
25
30
±(0.012 + .06/°C)
±(0.079 + 1.0/°C)
±(0.266 + .06/°C)
±(1.782 + 1.0/°C)
1-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720etHp 8719esHp 8722etHp 8720es8722es

Table of Contents