Error Messages
Error Messages in Alphabetical Order
CANNOT READ/WRITE HFS FILE SYSTEM
Error Number
The disk is being accessed by the analyzer and is found to contain an HFS
203
(hierarchical file system) or files nested within subdirectories. The analyzer does
not support HFS. Replace the disk medium with a LIF or DOS formatted disk
that does not contain files nested within subdirectories.
CAN'T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY
Error Number
Your sequence transfer to or from a disk could not be completed due to
127
insufficient memory.
CAUTION: AUX CHANNELS MEASURE S-PARAMETERS ONLY
Error Number
This message appears if you try to assign a non-S-parameter measurement to
216
an auxiliary channel and a full 2-port cal is performed on channels 1 and 2 with
both channels uncoupled.
CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED
Error Number
This message is displayed when correction is forced off due to a stimulus change
215
that is not compatible with the current calibration while an auxiliary channel is
enabled. The auxiliary channels are restored when correction is turned on by
pressing
CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND
Error Number
Your target value for the marker search function does not exist on the current
159
data trace.
CONTINUOUS SWITCHING NOT ALLOWED
Error Number
Your current measurement requires different power ranges on channel 1 and
10
channel 2. To protect the attenuator from undue mechanical wear, test set hold
will be activated. The "tsH" (test set hold) indicator in the left margin of the
display indicates that the inactive channel has been put in the sweep hold mode.
This message is also displayed if a mechanical switch test set is in use and
channels are measuring parameters that require the test set to switch
continuously, for example S11 on Channel 1 and S22 on Channel 2
COPY: device not responding; copy aborted
Error Number
The printer or plotter is not accepting data. Verify the cable connections, HP-IB
170
addresses, and otherwise ensure that the copy device is ready.
COPY OUTPUT COMPLETED
Information
The analyzer has completed outputting data to the printer or plotter. The
Message
analyzer can now accept another copy command.
6-6
CORRECTION on OFF