HP 8719ET Reference Manual page 72

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

HP 8719/20/22ET Specifications and Characteristics
Instrument Specifications
Table 2-8 Test Port Output
HP 8719ET/20ET/22ET Test Port Output
Description
Output Power
a
Level Accuracy
HP 8719ET/20ET
HP 8722ET
Maximum Leveled Power
HP 8719/20ET
HP 8722ET
0.05 to 20 GHz
20 to 40 GHz
c
Power Range
HP 8719/20ET (Standard)
HP 8719/20ET (Option 004)
HP 8722ET (Standard)
0.05 to 20 GHz
20 to 40 GHz
HP 8722ET (Option 004)
0.05 to 20 GHz
20 to 40 GHz
Power Sweep Range
HP 8719/20ET
HP 8722ET
0.05 to 20 GHz
20 to 40 GHz
a. Absolute power accuracy at a given power level. Includes absolute accuracy and relative
flatness across frequency.
b. At any given frequency, the achievable power while remaining leveled. Applies to CW
mode only.
c. Power to which the source can be set and phase lock is assured.
2-12
Specification
± 2 dB
± 3 dB
b
−10 to +10 dBm
−65 to +10 dBm
−15 to 0 dBm
−15 to −5 dBm
−70 to 0 dBm
−70 to−5 dBm
20 dB
15 dB
10 dB
Supplemental Information
at 5 dBm
at −5 dBm
+10 dBm, char.
0 dBm, char.
−5 dBm, char.
−18 to +12 dBm, typ.
−23 to +2 dBm, typ.
−23 to −3 dBm, typ.
30 dB, typ.
25 dB, typ.
20 dB, typ.