HP 8719ET Reference Manual page 33

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

Table 1-20 Type-N Device Connector Type
Network Analyzer: HP 8722ES, Standard
Calibration Kit: HP 85054B (Type-N, 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
Magnitude (dB)
Phase (deg)
Trans. Tracking
Magnitude (dB)
Phase (deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
HP 8719/20/22ES Specifications and Characteristics
0.05 to 2 GHz
48
46
48
±(0.005 + .03/°C)
±(0.033 + 0.1/°C)
±(0.013 + .03/°C)
±(0.086 + 0.1/°C)
Corrected System Performance (HP 8722ES)
Specification
2 to 8 GHz
42
36
42
±(0.006 + .03/°C)
±(0.040 + 0.3/°C)
±(0.041 + .03/°C)
±(0.271 + 0.3/°C)
8 to 18 GHz
42
33
42
±(0.015 + .04/°C)
±(0.099 + 0.5/°C)
±(0.071 + .04/°C)
±(0.469 + 0.5/°C)
1-23

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720etHp 8719esHp 8722etHp 8720es8722es

Table of Contents