HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Table 1-2 System Dynamic Range, All Device Connector Types
HP 8722ES, All Cal Kits, All Cables, 10 Hz IF BW
Description
System Transmission Dynamic Range
HP 8722ES (Standard)
50 MHz to 840 MHz
840 MHz to 8 GHz
8 GHz to 20 GHz
20 GHz to 40 GHz
HP 8722ES (Option 007)
50 MHz to 840 MHz
840 MHz to 8 GHz
8 GHz to 20 GHz
20 GHz to 40 GHz
HP 8722ES (Option 012 with test set bypassed)
50 MHz to 500 MHz
50 Hz to 8 GHz
8 GHz to 20 GHz
20 GHz to 40 GHz
a. The System Transmission Dynamic Range is calculated as the difference
between the receiver noise floor and the lesser of either: the source maximum
output or the receiver maximum input. The receiver noise floor is specified as 3
standard deviations above the mean of the linear magnitude noise floor trace
over the specified frequency band.
b. 3 dB less for Option 085 or Option 012.
c. The System Transmission Dynamic Range is calculated as the difference
between the receiver noise floor and the lesser of either: the source maximum
output or the receiver maximum input. The receiver noise floor is specified as
the mean of the linear magnitude noise floor trace over the specified frequency
band.
1-4
a
a
Specification
67 dB
93 dB
91 dB
b
80 dB
72 dB
98 dB
96 dB
b
85 dB
c
Supplemental
Information
115 dB, typ.
115 dB, typ.
113 dB, typ.
108 dB, typ.