Hardkey/Softkey Reference
Analyzer Functions
C0
C1
C2
C3
Cal
CAL FACTOR
CAL FACTOR SENSOR A
CAL INTERP ON off
CAL FACTOR SENSOR B
5-10
is used to enter the C0 term in the definition of an OPEN
standard in a calibration kit, which is the constant term of
the cubic polynomial and is scaled by 10
is used to enter the C1 term, expressed in F/Hz
(Farads/Hz) and scaled by 10
is used to enter the C2 term, expressed in F/Hz
−36
scaled by 10
.
is used to enter the C3 term, expressed in F/Hz
−45
scaled by 10
.
key leads to a series of menus to perform measurement
calibrations for vector error correction (accuracy
enhancement), and for specifying the calibration
standards used. The
which activate interpolated error correction and power
meter calibration.
accepts a power sensor calibration factor % for the
segment.
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a
power sensor's calibration factors. The calibration factor
data entered in this menu will be stored for power sensor
A.
sets the preset state of interpolated error-correction on or
off.
brings up the segment modify menu and segment edit
(calibration factor menu) which allows you to enter a
power sensor's calibration factors. The calibration factor
data entered in this menu will be stored for power
sensor B.
−15
−27
.
key also leads to softkeys
CAL
.
2
and
3
and