HP 8719ET Reference Manual page 30

Network analyzers
Hide thumbs Also See for HP 8719ET:
Table of Contents

Advertisement

HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8722ES)
Table 1-17 Type-N Device Connector Type
Network Analyzer: HP 8722ES, Standard
Calibration Kit: HP 85054D (Type-N, 50 Ω)
Cables: HP 85132F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
Magnitude (dB)
Phase (deg)
Trans. Tracking
Magnitude (dB)
Phase(deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
1-20
0.05 to 2 GHz
40
38
40
±(0.006 + .03/°C)
±(0.040 + 0.1/°C)
±(0.026 + .03/°C)
±(0.172 + 0.1/°C)
Specification
2 to 8 GHz
36
33
36
±(0.009 + .03/°C)
±(0.059 + 0.3/°C)
±(0.070 + .03/°C)
±(0.462 + 0.3/°C)
8 to 18 GHz
34
29
34
±(0.027 + .04/°C)
±(0.178 + 0.5/°C)
±(0.128 + .04/°C)
±(0.845 + 0.5/°C)