Table 1-3 3.5-mm Device Connector Type
Network Analyzer: HP 8719ES/20ES, Standard
Calibration Kit: HP 85052D (3.5-mm, 50 Ω)
Cables: HP 85131F
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Directivity (dB)
Source Match (dB)
Load Match (dB)
Refl. Tracking
Magnitude (dB)
Phase (deg)
Trans. Tracking
Magnitude (dB)
Phase (deg)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
50 to 500 MHz
0.5 to 2 GHz
42
37
42
±(0.006 + .02/°C)
±(0.006 + .03/°C)
±(0.040 + 0.1/°C)
±(0.040 + 0.1/°C)
±(0.028 + .02/°C)
±(0.03 + .03/°C)
±(0.185 + 0.1/°C)
±(0.198 + 0.1/°C)
HP 8719/20/22ES Specifications and Characteristics
Corrected System Performance (HP 8719/20ES)
Specification
2 to 8 GHz
42
38
37
31
42
38
±(0.006 + .03/°C)
±(0.040 + 0.3/°C)
±(0.096 + .03/°C)
±(0.634 + 0.3/°C)
8 to 20 GHz
36
28
36
±(0.009 + .04/°C)
±(0.059 + 0.5/°C)
±(0.158 + .04/°C)
±(1.04 + 0.5/°C)