6 Start-up
⇒
PARAM
CODE
T1
T2
T3
T4
T5
26
6.4
Quick Setup
In this menu, configure the most important transmitter functions required for
measurement.
The Quick Setup ist started automatically when starting the instrument. You can open
the Quick Setup at any time from the menu structure.
To enter the menu, proceed as follows:
DISPLAY
Hold
Language
Hold
+-
Hold
Date+time
Weekday
Mo
Day
30
Year
Time
Select
Hold
Sensor input
Hold
Meas.principle
2-circuit
⇒
CHOICE
(default = bold)
E
D
Mo
01
04
01
12:00
pH
Redox mV
Redox %
One circuit K1
One circuit K2
Two circuit
Mycom S CPM 153
⇒
INFO
Select language
Depending on ordered language version
Language version variants:
-A: E / D
-B: E / F
-C: E / I
-D: E / ES
-E: E / NL
-F: E / J
Contrast setting as necessary
You can increase and reduce the
contrast with the +/- keys.
Date and time entry
Enter the complete date and time here.
!
Operating mode selection
Note!
• If you change the operating mode, all
user settings are automatically reset!
• Here the use of the DAT module can be
helpful for saving your settings.
Measuring principle selection
(one-circuit K2 and two circuits only with
two-circuit instrument)
One-circuit K1 / K2 = measurement via
sensor input 1 or 2
Two-circuit = measurement via both sen-
sor inputs