Mycom S CPM 153
Endress+Hauser
4.2
Sensor connection and measuring cable
4.2.1
Cable types
You require a screened special cable to connect pH and redox electrodes. You can use
the following multicore and preterminated cable types:
• CPK 1 for electrodes without Pt 100
• CPK 6 for:
– Sensopac with double reference electrode
– 1 pH individual electrode, 1 reference electrode, 1 Pt 100
– 2 pH combination electrodes with 1 Pt 100 or
2 pH combination electrodes with 1 shared reference electrode
• CPK 9 for electrodes with TOP 68 plug-in heads (ESA / ESS) and Pt 100
• CPK 12 for IsFET pH sensors and pH/redox glass electrodes with TOP 68 plug-in
heads (ESB) and Pt 100 / Pt 1000
PA
CPK 1
pH / Redox
PC ye/gn
CPK 6-S
Application:
pH
Sensopac
Ref.
CPA 320 with
double reference
Pt 100
electrode
PA bk
CPK 6-D
PC
Application:
PA
1 pH single
pH
electrode,
1 reference
Ref.
electrode and
1 Pt 100
Pt 100
CPK 6-A
PE
Application:
PA
2 pH combination
pH 1
electrodes
and 1 Pt 100 or
pH 2
2 pH combination
electrodes
Pt 100
and 1 common
reference electrode
(alternative: reference electrode)
CPK 9
for pH/redox elec-
pH/
trodes with and without
Redox
integrated temperature
+Pt 100
sensor (for TOP 68
plug-in heads ESA/ESS)
CPK 12
for IsFET pH sensors
pH/
and pH/redox glass
Redox
electrodes with
+Pt 100/
and without integrated
Pt 1000
temperature sensor
(for TOP 68 plug-in heads
ESA/ESS)
Fig. 10:
Preterminated special measuring cables to connect pH and redox electrodes.
Outer screen
Outer screen
ye/gn
Ye/gn is connected
Outer screen: ground via metal cable union
bk
to outer screen
or connect to PE
rd
1
wt
2
gn
3
Under sleeve,
wt
wt connected to bn
Outer screen: ground via metal cable union
or connect to PE
ye/gn
Ye/gn is connected
bk
to outer screen
rd
1
wt
2
gn
Unter sleeve, wt
3
wt
connected to bn
PA
PA
4 Electrical connection
Ref.
Coax bk
pH / Redox
bn
PA
bk
PA
Coax rd
pH for circuit 1
wt
pH for circuit 2
Coax wt
Ref. 2
Ref. 1
wt
gn
Temp. (pin 11)
wt
Temp. (pin 12)
bn
Temp. (pin 13)
bk
PA
Coax rd
pH
wt
remove
Coax wt
Ref.
wt
Screen to 0 V (PA)
gn
Temp.
wt
Temp.
bn
Temp.
Slot 1
Signal
FCP1 (1)
rd
Ref. 1
Ref.
Remove from PA, attach end sleeve
PA
PA
bk
Coax rd
pH 1
pH/mV
remove
Coax wt
pH 2
- - -
Coax wt
Coax wt
Ref. 2
- - -
gn
Temp. (Ref.) 11 (Ref.) - - - (Ref.)
wt
Temp. (- - -) 12
bn
Temp. (- - -) 13
Ref.
Coax bk
pH
gn
wt
ye
d.n.c.
PA
bn
IsFET
Glass el.
source
pH
reference reference
Coax bk
rd
drain
gn
wt
ye
bn
PA
PA
Slot 2
FCP1 (2)
- - -
- - -
- - -
pH/mV
Ref.
- - -
- - -
15
Need help?
Do you have a question about the Mycom S CPM 153 and is the answer not in the manual?
Questions and answers