Endress+Hauser Mycom S CPM 153 Operating Instructions Manual page 96

Ph and redox transmitter
Table of Contents

Advertisement

8 Troubleshooting
Error
NAMUR
Error message
no.
class
E055
Failure
Display range of main
parameter 1 under-
shot
E056
Failure
Display range of main
parameter 2 under-
shot
E057
Failure
Display range of main
parameter 1
exceeded
E058
Failure
Display range of main
parameter 2
exceeded
E059
Failure
Temperature range 1
undershot
E060
Failure
Temperature range 2
undershot
E061
Failure
Temperature range 1
exceeded
E062
Failure
Temperature range 2
exceeded
E063
Mainte-
Current limit
nance
0/4 mA output 1
E064
Mainte-
Current limit 20 mA
nance
output 1
E065
Mainte-
Current limit 0/4 mA
nance
output 2
E066
Mainte-
Current limit 20 mA
nance
output 2
E067
Mainte-
Reference value
nance
exceeded controller
LS 1
E068
Mainte-
Reference value
nance
exceeded controller
LS 2
E069
Mainte-
Reference value
nance
exceeded controller
LS 3
E070
Mainte-
Reference value
nance
exceeded controller
LS 4
E071
Mainte-
Reference value
nance
exceeded controller
LS 5
96
Possible causes / measures
Measuring line broken,
sensor in air or air cushion in assembly,
Potential matching missing in symmetrical
measurement, static charging in media
with lowest conductivity
Temperature sensor defective;
Sensor line interrupted or short-circuited;
Wrong sensor type selected
Measured value outside specified current
range:
Check measured value for plausibility,
if necessary adjust current output assign-
ment 0/4 mA and/or 20 mA.
Dosing devices defective;
Chemical supply empty;
Measured value incorrect -> check for
plausibility and function;
Incorrect control direction set;
incorrect contact assigned;
Incorrect control function assigned
Mycom S CPM 153
Alarm
Error
contact
current
Fact
User
Fact
User
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
Endress+Hauser
Automatic
cleaning
start
Fact
User
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no

Hide quick links: