Mycom S CPM 153
Endress+Hauser
CODE
CHOICE
(default = bold)
MBB7
Relay:
Period:
t
min:
E
Sensor technology:
MC1
Look-ahead measurement:
pH circuit 1 = controller
pH circuit 2 = look-ahead
MC2
Control with:
pH value circuit 1
pH value circuit 2
MC3
L
:
B
L
:
S
L
:
E
MC4
Unit:
Unit:
4 mA value:
20 mA value:
INFO
n.c.
Alkali dosing: Relay selection
000.0 s
(for pulse length)
000.0 s
Description see above
Note in display:
(only look-ahead)
In Quick Setup one process with look-ahead pH
!
measurement was selected.
Note!
Control with look-ahead measurement is only
possible in conjunction with a flowmeter and a two-
circuit transmitter.
Electrical assignment:
(only redundancy)
Selection of with which measured value, control is
effective.
0.5 m
System arrangement
0.5 m
Enter electrode/dosing point distances:
1.5 m
L
: Distance between the controlling electrode and the
S
acid dosing point
: Distance between the controlling electrode and the
L
B
alkali dosing point
: Distance from the controlling electrode to the
L
E
look-ahead electrode
Remark on Fig. 24: Electrode 1 is the controlling elec-
trode, electrode 2 is the look-ahead electrode.
Two-sided pipe flow neutralisation (inline)
with forward-locking pH measurement
pH elec-
Flow
trode 2
meter
Q
Fig. 24:
Schematic diagram of two-sided control with
look-ahead pH measurement
m
Flow velocity flowmeter
s
Unit: Entry of the length and time unit for flow velocity
___
(e.g. m/s).
___
4 mA value: Enter minimum flow velocity value.
20 mA value: Enter maximum flow velocity value.
6 Start-up
Mycom 153
pH elec-
trode 1
F
Q
Alkali
Acid
L S
L B
L E
63