6 Start-up
PARAM
52
6.5.11
Set up 2 – Check
To enter the menu, proceed as follows:
CODE
CHOICE
(default = bold)
L1
SCS K1:
SCS Ref. K1:
SCS K2:
SCS Ref. K2:
L2
PCS K1:
INFO
off
Select SCS (= Sensor Check System) mode for meas-
light
uring circuits 1 (K1) and 2 (K2) for the two-circuit
off
instrument:
medium
SCS: Recognition of glass breakage
(off; Ref. =Reference electr.; Glass=electrode;
G+R= Electr.+Reference electr.)
SCS Ref.: Blockage recognition
!
(off, light, medium, heavy, very heavy blockage)
Note!
In an unsymmetrical connection (without PML) only the
glass electrode can be monitored.
off
PCS (= Process Check System) time
If the measuring signal does not change during the
entered time for ±0,02 pH / ±5mV / ±0,25%, an alarm is
signalled with error message E152.
!
Settable times: off, 1h, 2h, 4h.
Note!
An active PCS alarm signal will be deleted
automatically as soon as the sensor signal changes.