Mycom S CPM 153
Error
NAMUR
Error message
no.
class
E027
Failure
Compressed air fail-
ure
E 030
Failure
SCS fault reference
electrode 1
E 031
Failure
SCS fault reference
electrode 2
E 032
Failure
Outside set slope
range for sensor 1
E 033
Failure
Outside set zero point
for sensor 1
E 034
Failure
Outside set offset
range for sensor 1
E 035
Failure
Outside set slope
range for sensor 2
E 036
Failure
Outside set zero point
for sensor 2
E 037
Failure
Outside set offset
range for sensor 2
E038
Mainte-
Delta limit exceeded
nance
E040
Mainte-
SCC / electrode sta-
nance
tus of sensor 1 bad
E041
Mainte-
SCC / electrode sta-
nance
tus of sensor 2 bad
E043
Mainte-
Buffer difference
nance
channel 1 too small
E044
Mainte-
Meas. value channel
nance
1 unstable
E045
Failure
Calibration aborted
E048
Mainte-
Buffer difference
nance
channel 2 too small
E049
Mainte-
Meas. value channel
nance
2 unstable
E054
Mainte-
Dosage time alarm
nance
Endress+Hauser
Possible causes / measures
Pressure below permitted minimum
Reference impedance too high:
Check reference element and, if neces-
sary replace reference or combination
electrode
For IsFET sensor: leak current > 400 nA.
Replace sensor.
Sensor aged or defective;
Reference aged, defective or diaphragm
blocked;
Buffer solutions too old or contaminated;
PML not in the buffer solutions
Sensor aged or defective;
Reference aged, defective or diaphragm
blocked;
Buffer solutions too old or contaminated;
PML not in the buffer solutions
Difference between channel 1 and 2
measured values too high. Process to
inconsistant or sensor defective. Replace
sensor if necessary.
Check sensor, replace if necessary;
maybe clean, (glass membrane blocked
or run dry; diaphragm blocked).
Wrong buffer used;
Buffer entry incorrect; buffer self-recogni-
tion defective.
PAL missing; sensor too old; sensor some-
times dry; cable or plug defective.
Repeat calibration and renew buffer solu-
tion. Replace electrode if necessary.
Wrong buffer used;
Buffer entry incorrect; buffer self-recogni-
tion defective.
PAL missing; sensor too old; sensor some-
times dry; cable or plug defective.
Dosage time exceeded at total dosage.
Dosage interrupted, dosing agent empty
or process too inconsistant.
8 Troubleshooting
Alarm
Error
contact
current
Fact
User
Fact
User
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
Automatic
cleaning
start
Fact
User
no
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
–
no
95
Need help?
Do you have a question about the Mycom S CPM 153 and is the answer not in the manual?