6 Start-up
50
CODE
CHOICE
(default = bold)
JC8
threshold
length
Calibration timer:
JD1
Cal timer: on
Warning: 0001h
___________
Time: 0001:00
INFO
02 mV
Stability
010s
During calibration the mV values may maximally vary for
the given threshold within the defined time range (length).
During calibration, the mV value may change during the
given period ("duration") at maximum by the stated
amount ("threshold"), so that the calibration is
considered as stable.
Therefore you can adjust accuracy and timing
individually to your process.
Calibration timer
If no calibration is undertaken in the set time, an error
message appears (E115).
Cal Timer: on = activate
Warning: Enter the time within which a calibration must
take place.
Time: Display of the remaining time to an error message.
Mycom S CPM 153
Endress+Hauser