Mycom S CPM 153
⇒
Endress+Hauser
6.5.17
Diagnosis
To enter the menu, proceed as follows:
CODE
CHOICE
(Default = bold)
U
Error list
Error log
Operation log
Calibration log
Service
Service:
Y
Factory settings
Simulations
Check systems
Reset
DAT handling
Instrument version
Factory function
Factory settings:
YA1
Cancel
Settling data
Calibration data
All data
Service data
Operation logbook
Error logbook
Calibration logbook
⇒
INFO
Error list: Displays the current active errors.
(Complete error list with description s. page 93)
Error log: Lists the last 30 signalled errors with date and
time.
Operation log (service code necessary): Lists the last
30 registered operating steps with date and time.
Calibration log: Lists the last 30 calibrations undertaken
!
with date and time.
Note!
• Use the arrow keys to scroll through the lists.
• Leave the lists by pressing "E".
Select service diagnosis
Factory settings: Different data groups can be reset to
the factory settings.
Simulations: The transmitter behaviour can be simula-
ted after entering of different parameters.
Check systems: The instrument functions (display,
keys, etc.) can be tested individually.
Reset: Device reset (no data loss!)
DAT handling: Copy data into/out of the DAT module.
Instrument version: Device-internal data e.g. serial
number can be queried.
Factory function: reset counter, write access
Set default
Here you can select the data which you wish to
!
reset to the factory settings.
Note!
Danger of data loss. Selecting a point and confirming
with "Enter" deletes all the settings you made in this
area!
Pressing Cancel leaves this field without changing the
values.
Calibration data: All the saved data for calibrations
such as zero point, slope, and offset.
Setting data: the remaining data to be set.
All data: calibration data + setting data
Service data: all data + logbooks + reset counters.
Service data / logbooks:
these functions are only for authorised service
personnel. The service code is required.