Chapter 6
RDO/DO Measurement Module
86
Orion Versa Star Pro Meter User Manual
RDO and DO Mode Menu
Main Menu
Settings
Electrode SI No
----
Sample ID
Off, Manual, Auto Increment
Measurement Unit
mg/L,
Resolution
0.1, 0.01
Stability
Smart, Fast, Medium, Slow
Averaging
Off, Automatic Smart
Auto-Read, Timed,
Read Type
Single-Shot, Continuous
Auto, Manual
Pressure Comp
775.0
Salinity Correction
Auto, Manual
Limit Alarm
Alarm
Cal Due Alarm
Set Point Alarm Off, On
Detailed information on these features is available in the

RDO Electrode Information

The serial number, remaining cap life and date of manufacture for each RDO optical cap can be
viewed in the RDO Electrode Information menu when the RDO probe is connected to the meter
and recognized in the measurement mode.
Verify that the RDO probe is connected to the meter. In the measurement mode, press the
setup key. Press the  or  key to highlight Channel # (channel with the RDO probe) and
press the f3 (Select) key. Press the  or  key to highlight RDO Electrode Information and
press the f3 (Select) key.
Details
Use the alphanumeric keypad to enter the serial
number or clear all to shown no value (----)
Use the popup keypad screen to enter an
alphanumeric manual ID or a numeric value as
the starting value for auto incremental ID
%
Set the displayed measurement units
1, 0.1
Set the resolution of the measurement value
Set when a measurement is recognized as
stable
Set averaging for faster measurement stability
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Set the barometric pressure compensation
source as Auto to use the built-in barometer or
Manual to enter a value. Select the Auto option
to calibrate the barometer pressure reading
Set the salinity correction source as Auto to use
a conductivity module and probe to measure the
sample salinity or Manual to enter a value
Off, On
Set high and low limit values to activate alarm
Off, On
Set a calibration interval (hours) to activate alarm
Set a base and offset value to activate alarm
Appendix
section of this user manual.