a. Follow all electrode rinsing and drying instructions or exceptions in the ISE user
manual. Do not wipe or rub the sensing element of the electrode.
13. Wait for the mV value on the meter to stabilize and press the f2 (Accept) key.
14. Enter the analate/sample volume using the numeric keypad and press the f2 (Accept) key.
15. Add the analate/sample to the standard and press the f2 (Yes) key.
16. Wait for the mV value on the meter to stabilize and press the f2 (Accept) key.
17. The meter will display the calculated sample concentration and summary of the
incremental technique performed. Press the f1 (Esc) key to proceed to the measurement
mode or press the f2 (Next) key to perform another analysis.
Saving the Incremental Technique as a Method
For frequently used incremental techniques, Versa Star Pro meters now allow individual
incremental techniques to be stored as methods.
1. In the measurement mode, press the setup key.
2. Press the ,,  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4
and press the f3 (Select) key.
3. Press the  or  key to highlight Method and press the f3 (Select) key.
4. Press the  or  key to highlight Edit / Create and press the f3 (Select) key.
5. Press the  or  key to highlight a new method and press the f3 (Edit) key.
6. Press the  or  key to highlight Inc. Tech and press the f3 (Select) key.
7. Press the  or  key to highlight Single Analate Addition and press the f3 (Select) key.
8. Press the  or  key to highlight a value and press the f3 (Edit) key. Use the meter's
numeric keypad to enter the value and press the f2 (Accept) key. Repeat for all values.
9. Once all the values for Single Analate Addition have been entered, press the f2 (Save)
key.
10. The meter will display the ISE measure mode setup menu. Review and update the
measurement parameters for the method and then press the f1 (Back) key.
11. Press the  or  key to highlight Load from the main method menu and press the f3
(Select) key.
12. Press the  or  key to highlight the ISE method for Single Analate Addition and press
the f3 (Load) key. The meter will return to the measurement mode with the method loaded.
Chapter 4