Conductivity, Tds, Salinity And Resistivity Mode Settings; Cell Constant (K) And Automatic Conductivity Calibration - Thermo Scientific Orion Versa Star Pro User Manual

Electrochemistry meters
Table of Contents

Advertisement

Appendix A
Channel Specific Settings and Features
140
Orion Versa Star Pro Meter User Manual
Conductivity, TDS, Salinity and Resistivity
Mode Settings

Cell Constant (K) and Automatic Conductivity Calibration

Versa Star Pro meters with a conductivity module are capable of automatically recognizing
Thermo Scientific Orion 100 μS/cm conductivity standard, 1413 μS/cm conductivity standard
and 12.9 mS/cm conductivity standard when the nominal cell constant (K) value of the
conductivity probe is entered in the setup menu.
Catalog Number
011008
Orion 100 μS/cm conductivity standard, 5 x 60 mL
011007
Orion 1413 μS/cm conductivity standard, 5 x 60 mL
01100710
Orion 1413 μS/cm conductivity standard, 10 individual use pouches
011006
Orion 12.9 mS/cm conductivity standard, 5 x 60 mL
01100610
Orion 12.9 mS/cm conductivity standard, 10 individual use pouches
The nominal cell constant (K) values for Thermo Scientific Orion conductivity probes compatible
with Versa Star Pro meters with a conductivity module are listed below.
Catalog
Number
Orion DuraProbe 4-cell conductivity probe with built-
013005MD
in temperature, 1.5 meter cable
Orion DuraProbe 4-cell conductivity probe with built-
013010MD
in temperature, 3 meter cable
Orion DuraProbe 4-cell conductivity probe with built-
013020MD
in temperature, 6 meter cable
Orion DuraProbe 4-cell conductivity probe with built-
013025MD
in temperature, 10 meter cable
Orion DuraProbe 4-cell conductivity probe with built-
013605MD
in temperature, 1.5 meter cable
Orion DuraProbe 4-cell conductivity probe with built-
013610MD
in temperature, 3 meter cable
Orion ultra-pure water 2-cell conductivity probe with
013016MD
built-in temperature, 1.5 meter cable
Orion 2-cell conductivity probe with built-in
011510MD
temperature, 3 meter cable
Orion 2-cell conductivity probe with built-in
011050MD
temperature, 1.5 meter cable
Orion high range 2-cell conductivity probe, 1.5 meter
018020MD
cable
Description
Description
Measurement
Nominal Cell
Range
Constant
1 μS/cm to
K = 0.475 cm
200 mS/cm
1 μS/cm to
K = 0.475 cm
200 mS/cm
1 μS/cm to
K = 0.475 cm
200 mS/cm
1 μS/cm to
K = 0.475 cm
200 mS/cm
10 μS/cm to
K = 0.55 cm
200 mS/cm
10 μS/cm to
K = 0.55 cm
200 mS/cm
0.01 μS/cm to
K = 0.1 cm
-1
300 μS/cm
10 μS/cm to
K = 1.0 cm
-1
200 mS/cm
1 μS/cm to
K = 1.0 cm
-1
20 mS/cm
10 μS/cm to
K = 10 cm
-1
2000 mS/cm
Thermo Scientific
-1
-1
-1
-1
-1
-1

Hide quick links: