Method Menu - Thermo Scientific Orion Versa Star Pro User Manual

Electrochemistry meters
Table of Contents

Advertisement

Chapter 3
Meter Setup Menus
26
Orion Versa Star Pro Meter User Manual

Method Menu

Up to ten methods can be saved per channel in the Method menu within the Channel 1,
Channel 2, Channel 3 and Channel 4 setup menus for fast and easy recollection of a specific
measure mode, settings and setup parameters. The default method (M100, M200, M300 and
M400) is updated whenever the measure mode and channel setup parameters are changed.
When a custom method is created, loaded and active in the measurement mode and a
calibration is performed, that calibration will be saved to the method so every time the method is
loaded, the corresponding calibration will also be loaded. This feature is helpful when using two
or more electrodes on one channel. For example, a pH electrode and an ion selective electrode
(ISE) used on the same VSTAR-ISE pH/ISE module or a low range conductivity probe and a
standard range conductivity probe used on the same VSTAR-CND conductivity module.
Edit / Create
1. In the measurement mode, press the setup key.
2. Press the ,,  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4
and press the f3 (Select) key.
3. Press the  or  key to highlight Method and press the f3 (Select) key.
4. Press the  or  key to highlight Edit / Create and press the f3 (Select) key.
5. Press the  or  key to highlight an existing method to edit or a new method to create
and press the f3 (Edit) key.
6. Press the  or  key to highlight a desired measure mode and press the f3 (Select) key.
7. The meter will display the appropriate setup menu for the selected measure mode. Review
and update the measurement parameters for the method and then press the f1 (Back) key.
a. The displayed measure modes are dependent on the module that is connected to the
selected channel and the displayed measurement setup parameters are dependent on
the selected measure mode. Refer to the individual module chapters for detailed
information and instructions.
8. Press the  or  key to highlight Load from the main methods menu and load the custom
method or press the measure (esc) key to return to the measurement mode.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents