Conductivity Mode Menu
Main Menu
Settings
Auto-Read, Timed,
Read Type
Single-Shot, Continuous
Cell K
0.475
Ref. Temperature
5°C, 10°C, 15°C, 20°C, 25°C
Temp.
Linear, nLFn, nLFu, EP, Off
Compensation
Temperature Coeff.
2.10
Stability
Smart, Fast, Medium, Slow
Averaging
Off, Automatic Smart
Limit Alarm
Alarm
Cal Due Alarm
Set Point Alarm Off, On
Cell Type
Standard, USP
Electrode SI No
----
Sample ID
Off, Manual, Auto Increment
Detailed information on these features is available in the
Chapter 5
Details
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Enter the nominal cell constant (K) value of the
conductivity probe, value is used when
performing an automatic conductivity calibration
Set the reference temperature that all displayed
conductivity measurements will be reported to
using the selected temperature compensation
Set the temperature compensation type used to
calculate and report conductivity measurements
at the selected reference temperature
Enter the temperature coefficient value for linear
temperature compensation
Set when a measurement is recognized as
stable
Set averaging for faster measurement stability
Off, On
Set high and low limit values to activate alarm
Off, On
Set a calibration interval (hours) to activate alarm
Set a base and offset value to activate alarm
Select Standard for most conductivity probes or
USP for ultra-pure water 2-cell conductivity probe
(Catalog Number 013016MD)
Use the alphanumeric keypad to enter the serial
number or clear all to shown no value (----)
Use the popup keypad screen to enter an
alphanumeric manual ID or a numeric value as
the starting value for auto incremental ID
Appendix
Orion Versa Star Pro Meter User Manual
Conductivity Measurement Module
section of this user manual.