Thermo Scientific Orion Versa Star Pro User Manual page 118

Electrochemistry meters
Table of Contents

Advertisement

Chapter 9
Customer Services
118
Orion Versa Star Pro Meter User Manual
Orion Versa Star pH/ISE Module Specifications
Catalog Number
Range
Resolution
Relative Accuracy
pH
Calibration Points
Calibration Editing
Adjustable ISO Point
Input Impedance
Range
Resolution
mV / RmV
Relative Accuracy
Relative mV Mode
E
ORP Mode
H
Range
Resolution
Relative Accuracy
Units
Calibration Points
ISE
Calibration Editing
Adjustable ISO Point
Advanced Features
Incremental
Techniques
Range
Resolution
Temperature
Relative Accuracy
Offset Calibration
Source Options
BNC
Inputs
Pin Tip
8 pin MiniDIN
VSTAR-ISE
- 2.000 to 20.000
0.1, 0.01, 0.001
± 0.002
Up to 6
Yes
Yes
> 10^
ohms
12
± 2000.0 mV
0.1 mV
± 0.2 mV or ± 0.05 % of reading, whichever is greater
Yes
Yes
0.0001 to 19990
0.0001 minimum, 1 to 4 significant digits (user selectable)
± 0.2 mV or ± 0.05 % of reading, whichever is greater
ppm, Molar, mg/L, %, ppb, none
Up to 6
Yes
Yes
Segmented (point-to-point) slope or linear regression, non-
linear selectable auto-blank, low concentration range
stability, electrode type display, temperature compensation
Single known addition & subtraction, double known addition
& subtraction, single analate addition & subtraction
-5 to 105 °C, 22 to 221 °F
0.1 °C, 0.1 °F
± 0.1 °C, ± 0.1 °F
1 point
Manual or automatic with ATC probe
pH, ORP or ion selective electrode (ISE)
Reference electrode
ATC temperature probe
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents