Temperature Menu

Temperature settings can be updated in the Temperature menus within the Channel 1, Channel
2, Channel 3 and Channel 4 setup menus. ATC (Automatic Temperature Compensation)
probes are defined as both separate temperature probes and those built into electrodes.
Temperature Unit
Versa Star Pro meters are capable of displaying the temperature units as degrees Celsius (°C)
or degrees Fahrenheit (°F).
1. In the measurement mode, press the setup key.
2. Press the ,,  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4
and press the f3 (Select) key.
3. Press the  or  key to highlight Temperature and press the f3 (Select) key.
4. Press the  or  key to highlight Temperature Unit and press the  or  key to
highlight °C or °F as the displayed temperature units.
5. Press the measure (esc) key to return to the measurement mode.
Temperature Input
The temperature source for each channel can be set to any available ATC probe connected to
the meter. Only channels with an ATC probe actively connected will be shown and selectable.
The temperature source can also be set to manual (MAN) using this menu.
One ATC probe can be used for multiple channels. For example, when a pH electrode and
conductivity probe with built-in temperature sensor are measuring the same sample, the pH
electrode on channel 1 can use a conductivity probe on channel 2 as the temperature source,
so a second ATC probe is not needed.
1. In the measurement mode, press the setup key.
2. Press the ,,  or  key to highlight Channel 1, Channel 2, Channel 3 or Channel 4
and press the f3 (Select) key.
3. Press the  or  key to highlight Temperature and press the f3 (Select) key.