Data Transfer Examples
5 Point pH Calibration, Printer Format
5 Point pH Calibration, CSV Format
HyperTerminal View:
Text:
VERSA STAR,V00000,2.73,Ch-1,
SENSOR-01,VA00000,M-101,02-Aug-12,3:40 PM,
USER-01,5,1,pH,2,mV,290.5,25.0,C (MAN),
Auto&Manual,2,pH,4.01,mV,174.3,25.0,C (MAN),
Auto&Manual,3,pH,6.997,mV,0.0,25.0,C (MAN),
Auto&Manual,4,pH,10.013,mV,-173.7,25.0,
C(MAN),Auto&Manual,5,pH,12,mV,-289.5,25.0,C
(MAN),Auto&Manual,1,-57.8,mV,2,-58.4,mV,3,
-57.6,mV,4,-58.3,mV,1,1.4,mV,2,-0.2,mV,3,-0.2,
mV,4,2.0,mV,98.1,%,103
Chapter 8
Data Transfer and Software Updates
pH Measurement, Printer Format
pH Measurement, CSV Format
HyperTerminal View:
Text:
VERSA STAR,V00000,2.73,02-Aug-12, 4:03:40
PM,Ch-1,SENSOR-01,3,USER-01,
VA00000,PH,6.997,pH,0.0,mV,25.0,C
(MAN),98.1, %,101