Orion Conductivity Standard Values From 0 °C To 50 °C - Thermo Scientific Orion Versa Star Pro User Manual

Electrochemistry meters
Table of Contents

Advertisement

Thermo Scientific
Orion Conductivity Standard Values from 0 °C to 50 °C
Orion
111.9mS/cm
Conductivity
Standard
Temperature
(°C)
Cat. No.
011005,
01100510
(mS/cm)
0
65.10
1
66.84
2
68.59
3
70.35
4
72.12
5
73.91
6
75.70
7
77.50
8
79.32
9
81.15
10
82.98
11
84.83
12
86.69
13
88.56
14
90.45
15
92.34
16
94.24
17
96.15
18
98.08
19
100.0
20
102.0
21
103.9
22
105.9
23
107.9
24
109.9
25
111.9
Appendix A
Orion
Orion
12.9mS/cm
1413μS/cm
Conductivity
Conductivity
Standard
Standard
Cat. No.
Cat. No.
011006,
011007,
01100610
01100710
(mS/cm)
(μS/cm)
7.135
776
7.344
799
7.555
822
7.768
846
7.983
870
8.200
894
8.418
918
8.638
943
8.860
968
9.084
992
9.309
1017
9.535
1043
9.763
1068
9.993
1094
10.22
1119
10.46
1145
10.69
1171
10.93
1198
11.16
1224
11.40
1251
11.64
1277
11.88
1304
12.12
1331
12.36
1358
12.61
1386
12.85
1413
Orion Versa Star Pro Meter User Manual
Channel Specific Settings and Features
Orion
Orion
147μS/cm
100μS/cm
Conductivity
Conductivity
Standard
Standard
Cat. No.
Cat. No.
01100910
011008
(μS/cm)
(μS/cm)
81
54
83
56
86
58
88
59
91
61
93
63
96
64
98
66
101
68
103
70
106
72
108
73
111
75
114
77
116
79
119
81
122
83
125
85
127
87
130
88
133
90
136
92
138
94
141
96
144
98
147
100
141