Thermo Scientific Orion Versa Star Pro User Manual page 79

Electrochemistry meters
Table of Contents

Advertisement

Thermo Scientific
Salinity Mode Menu
Main Menu
Settings
Auto-Read, Timed,
Read Type
Single-Shot, Continuous
Cell K
0.475
Salinity Type
Practical Salinity, Sea Water
Stability
Smart, Fast, Medium, Slow
Averaging
Off, Automatic Smart
Limit Alarm
Alarm
Set Point Alarm Off, On
Cell Type
Standard, USP
Detailed information on these features is available in the
Resistivity Mode Menu
Main Menu
Settings
Auto-Read, Timed,
Read Type
Single-Shot, Continuous
Cell K
0.475
Ref. Temperature
5°C, 10°C, 15°C, 20°C, 25°C
Temp.
Linear, nLFn, nLFu, EP, Off
Compensation
Temperature Coeff.
2.10
Stability
Smart, Fast, Medium, Slow
Averaging
Off, Automatic Smart
Limit Alarm
Alarm
Set Point Alarm Off, On
Cell Type
Standard, USP
Detailed information on these features is available in the
Chapter 5
Details
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Enter the nominal cell constant (K) value of the
conductivity probe; value is used when
performing an automatic conductivity calibration
Set the salinity type used to convert conductivity
to salinity measurements
Set when a measurement is recognized as
stable
Set averaging for faster measurement stability
Off, On
Set high and low limit values to activate alarm
Set a base and offset value to activate alarm
Select Standard for most conductivity probes or
USP for ultra-pure water 2-cell conductivity probe
Appendix
Details
For Timed and Single-Shot, set the time in hours,
minutes, seconds using the arrow keys
Enter the nominal cell constant (K) value of the
conductivity probe; value is used when
performing an automatic conductivity calibration
Set the reference temperature that all displayed
resistivity measurements will be reported to using
the selected temperature compensation
Set the temperature compensation type used to
calculate and report resistivity measurements at
the selected reference temperature
Enter the temperature coefficient value for linear
temperature compensation
Set when a measurement is recognized as stable
Set averaging for faster measurement stability
Off, On
Set high and low limit values to activate alarm
Set a base and offset value to activate alarm
Select Standard for most conductivity probes or
USP for ultra-pure water 2-cell conductivity probe
Appendix
Orion Versa Star Pro Meter User Manual
Conductivity Measurement Module
section of this user manual.
section of this user manual.
79

Hide quick links:

Advertisement

Table of Contents
loading