Single-Shot; Continuous - Thermo Scientific Orion Versa Star Pro User Manual

Electrochemistry meters
Table of Contents

Advertisement

Thermo Scientific

Single-Shot

When Single-Shot is selected as the read type and a wait time is entered, a single
measurement is recorded, locked and held on the display after the predefined wait time has
elapsed. To start a new measurement using the same wait time, press the measure (esc) key.
Depending on the meter communication settings in the
recorded measurement value will be exported to the data log, printer and/or computer.

Continuous

When Continuous is selected as the read type, measurements are continuously updated on the
display and the Stabilizing and Ready icons specify the measurement value stability. This is
useful when performing an experiment that requires continuous measurements to be observed.
Depending on the meter communication settings in the
the log/print key to export a measurement to the data log, printer and/or computer.
Read Type Priority
When two or more modules are connected to the meter and displayed in the measurement
mode, the Versa Star Pro meter will use a priority structure to determine when the displayed
measurements are sent to the data log and external devices, if they are enabled in the
Instrument Settings
setup menu.
Primary
Additional
Priority
Read Type
Read Types
Timed
Auto-Read
1
Timed
Single-Shot
Continuous
Auto-Read
2
Auto-Read
Single-Shot
Continuous
Single-Shot
3
Single-Shot
Continuous
4
Continuous Continuous
Instrument Settings
Instrument Settings
Displayed Measurements
Continuously updated
Locked when stable, press the
measure (esc) key to update
Locked when wait time elapses, press
the measure (esc) key to update
Continuously updated
Locked when stable, press the
measure (esc) key to update
Locked when wait time elapses or all
Auto-Read channels are stable,
whichever occurs first
Continuously updated
Locked when wait time elapses, press
the measure (esc) key to update
Continuously updated
Continuously updated
Orion Versa Star Pro Meter User Manual
Chapter 7
Meter Measurements
setup menu, the
setup menu, press
Export to Data
Log and/or Device
All displayed
measurements are
exported at shortest
time interval setting
All displayed
measurements are
exported when all
Auto-Read channels
are stable
All displayed
measurements are
exported when wait
time elapses for all
Single-Shot channels
Press the log/print
key to exported all
displayed
measurements
93

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents