Conductivity Calibration Editing
Once the f3 (Cal Done) key is pressed and the calibration is saved, press the f2 (Cal Edit) key
to edit the calibration data and fix individual points without a full recalibration. In the calibration
edit mode, press the  or  key to highlight a calibration point and press the f2 (Delete) key
to delete the point or press the f3 (Re-Measure) key to re-measure the point. Follow the
onscreen text prompts and instructions to edit the selected calibration point.
Orion Conductivity Standard Values from 15 °C to 30 °C
Refer to the
Appendix
section of this user manual for additional values from 0 to 50°C.
Orion
111.9mS/cm
Standard
Temperature
(°C)
Cat. No. 011005,
01100510
(mS/cm)
15
92.34
16
94.24
17
96.15
18
98.08
19
100.0
20
102.0
21
103.9
22
105.9
23
107.9
24
109.9
25
111.9
26
113.9
27
115.9
28
117.9
29
120.0
30
122.0
Chapter 5
Orion
Orion
12.9mS/cm
1413μS/cm
Standard
Standard
Cat. No.
Cat. No.
011006,
011007,
01100610
01100710
(mS/cm)
(μS/cm)
10.46
1145
10.69
1171
10.93
1198
11.16
1224
11.40
1251
11.64
1277
11.88
1304
12.12
1331
12.36
1358
12.61
1386
12.85
1413
13.10
1441
13.35
1468
13.59
1496
13.84
1524
14.09
1552
Orion Versa Star Pro Meter User Manual
Conductivity Measurement Module
Orion
Orion
147μS/cm
100μS/cm
Standard
Standard
Cat. No.
Cat. No.
01100910
011008
(μS/cm)
(μS/cm)
119
81
122
83
125
85
127
87
130
88
133
90
136
92
138
94
141
96
144
98
147
100
150
102
153
104
156
106
159
108
161
110
81