Chapter 3
Meter Setup Menus
34
Orion Versa Star Pro Meter User Manual

Log View Menu

Use the Log View menu to access the data log and calibration log. All information in the data
and calibration logs is protected by the nonvolatile memory, so stored information is saved in
the meter even when it is not connected to a power supply.

Data Log

Versa Star Pro meters offer a 2000 point data log. Each point includes measurements from one
to four channels, depending on which channels were actively displayed in the measurement
mode when the point was saved. When the data log function is turned on, the read type
selected for each displayed channel (Auto-Read, Timed, Single-Shot or Continuous)
determines how the point is saved to the data log. Refer to the
section for detailed information sending measurements to the data log.
1. In the measurement mode, press the setup key.
2. Press the ,,  or  key to highlight Log View and press the f3 (Select) key.
3. Press the  or  key to highlight Data Log.
4. To view the log, press the  or  key to highlight View and press the f3 (Select) key. A
list of the saved data log points will be displayed with the data log numbers, main
measurement value of each channel, date and time.
a. Press the  or  key to highlight a point and press the enter key to view detailed
data for that point.
b. When viewing individual data log points, press the  or  key to scroll through
additional data log points.
5. To print the log, press the  or  key to highlight Print single point, Print range or Print all
and press the f3 (Select) key.
a. For a single point or range, enter the data point values and press the f3 (Print) key.
6. To delete the log, press the  or  key to highlight Delete all and press the f3 (Select)
key. Press the f1 (Yes) key to delete the log or press the f3 (No) key to cancel the function.
7. To view statistics, press the  or  key to highlight Statistics and press the f3 (Select)
key. Press the  or  key to select a measure mode and the statistical values for all data
log points with the selected measure mode will be displayed.
Measurement Read Types
Thermo Scientific