Chapter 8
Data Transfer and Software Updates
108
Orion Versa Star Pro Meter User Manual
Mode
Format Example
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, RDO
Concentration Value, RDO Concentration Unit, RDO Saturation Value, RDO Saturation
Unit, Partial Pressure in Torr, Solution Temperature Value, Solution Temperature Unit,
Barometric Pressure Value, Barometric Pressure Unit, Salinity Correction Value in ppt,
RDO mg/L
Slope Value, Slope Unit, Method Number
Example: Versa Star Pro, V11702, 11.20, 29-03-18, 3:50:53 PM, Ch-4, SENSOR-04, 1,
ADMIN, VA12345, RDO, 8.59, mg/L, 102.0, %, 150.052, 24.3, C (ATC), 760.2, mmHg,
0.1, 1.6, Torr/%Sat, M401 <CR>
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time
Multiple
Channels
followed by the measurement data for each channel as shown above