Channel Specific Settings and

Appendix A

Features
Within the Channel 1, Channel 2, Channel 3 and Channel 4 setup menus is the Mode menu,
which can be used to customize measurement settings and parameters for the selected
channel. This section provides detailed information on the options available within Mode menus
when using the various Versa Star measurement modules. Detailed information on additional
advanced meter features is also provided.

General Mode Settings

Sample ID

If Manual is selected as the type of Sample ID, use the popup alphanumeric keypad screen to
enter a sample ID up to ten characters long. The entered sample ID will be recorded with
subsequent measurements and saved to the data log or exported to an external device,
depending on the meter communication settings in the
Instrument Settings
setup menu.
If Auto Incremental is selected as the type of Sample ID, use the popup numeric keypad screen
to enter a numeric value up to ten characters long. The entered sample ID will be used as the
starting value and the meter will automatically increase the value by one for each subsequent
measurement. The sample IDs will be recorded with each measurement and saved to the data
log or exported to an external device, depending on the meter communication settings in the
Instrument Settings
setup menu.
Press the f2 (Sample ID) key in the measurement mode to open a shortcut to the Sample ID
setup menu and enter a new sample ID.
128
Orion Versa Star Pro Meter User Manual
Thermo Scientific