3 Prepare the e2695 Separations Module for Operation
Prepare the solvent management system for operation:
Start
Prepare solvent
reservoirs
Degas solvents
To maintain the efficiency of the solvent management system, and to
Tip:
obtain accurate, reproducible chromatograms, use only chromatography-grade
solvents.
Warning:
Refer to the Material Safety Data Sheets for the solvents you use.
Prepare solvent reservoirs
General recommendations:
•
When using the instrument for general chromatography (i.e., reverse
and/or normal phase and gel permeation [GPC]), Waters recommends
the use of high-quality lab glassware made of borosilicate glass for all
reservoirs (solvent, seal wash & needle wash).
•
When using techniques such as ion chromatography, where glass
containers can contribute ionic contamination (sodium and/or chloride
ions), it is acceptable to use laboratory-grade polypropylene or
polyethylene containers as reservoirs.
•
When using the instrument in combination with a mass spectrometric
detector, please refer to recommendations within the most recent
104
July 29, 2013, 715003794 Rev. B
Prime the plunger
seal-wash pump
Prime the solvent
management
system
Observe safe laboratory practices when you handle solvents.
Solvent
management
system is ready to
operate