Waters Alliance e2695 Operator's Manual page 93

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Auto shutdown
You can set up the separations module to shut down automatically after a
specified period of inactivity, which is defined as follows:
No keyboard use
No injections performed
No changes sent to the separations module from a remotely connected
Empower system, MassLynx computer, or external autosampler
An error condition that suspends the operation of the separations
module
You can leave the separations module in the shutdown state indefinitely. The
separation method you specify in the Auto Shutdown dialog box defines the
initial conditions that are applied after the specified period of inactivity. Use
the auto shutdown function:
When there is a long delay between injections
To minimize solvent flow after an unattended or long run
To disable the in-line vacuum degasser
To disable temperature controls
To turn off detector lamps
For instrument storage lasting one week or more, use methanol or
Tip:
methanol mixed with HPLC-quality water to flush the solvent delivery system
before shutdown (see
To enable the auto shutdown function:
1.
Press the Config screen key in the Main screen, and then press the More
screen key once.
2.
Press the Auto Shutdown screen key.
3.
Select a separation method to use while the separations module is shut
down.
Only the initial conditions in the method you select are used.
Tip:
4.
Enter a time period (in minutes) after which you want the separations
module to shut down.
Press Clear to disable the Auto Shutdown function.
Alternative:
5.
Press OK.
"Power-off the separations module" on page
Configure the separations module
July 29, 2013, 715003794 Rev. B
114).
93

Advertisement

Table of Contents
loading

Table of Contents