Load Carousels - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

A "Compression Check Failed" error message appears.
A "Missing Restrictor" alarm appears during a diagnostic procedure.
You change the seal pack or needle.
You rebuild the seal pack.
Before you adjust the seal pack, purge the sample management system to
ensure the syringe is empty of air. Air in the system can cause the Alarm Seal
Geometry or Alarm Missing Restrictor dialog boxes to (erroneously) appear.
To adjust the seal pack:
1.
In the Main screen, press the Diag screen key.
2.
In the Diagnostics screen, press the Adjust Seals screen key.
Result:
completes, the results are shown (see

Load carousels

The following table lists the carousels and their color codes.
Carousel codes and colors:
Code
Color
A
Blue
B
Yellow
C
Red
D
Green
E
White
To load the carousels:
1.
Open the carousel door.
The message the "Door is Open" appears.
Tip:
2.
Press the Next screen key (or select the desired carousel screen key) to
position the carousel turntable for loading the appropriate carousel.
3.
Load the carousel into the sample compartment.
4.
Repeat
Prepare the sample management system for operation
The adjustment procedure begins. When the procedure
step 2
and
step 3
until all carousels are loaded.
"Adjust the seals" on page
Vial numbers
1 to 24
25 to 48
49 to 72
73 to 96
97 to 120
July 29, 2013, 715003794 Rev. B
226).
111