Waters Alliance e2695 Operator's Manual page 155

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Alarm responses: (Continued)
Alarm response
Alert User
Stop Funct
Stop Flow
Set Gradient table parameter values
The Gradient table allows you to make time-based changes to the composition
of the mobile phase during a run. You can program up to 25 lines in the
Gradient table.
To set the parameters in the Gradient table:
1.
On the Mobile Phase screen, press the Gradient screen key to display
the Gradient screen.
Gradient screen:
Function
The error is entered into the error log and you are
alerted with a dialog box.
The error is entered into the error log, you are alerted
with a dialog box, and the operation is suspended at
the end of the current function. You can abort or
resume the operation of the sample set by pressing the
appropriate screen key.
The error is entered into the error log, you are alerted
with a dialog box, the current function is suspended,
and solvent flow is stopped.
Set separation method parameters
July 29, 2013, 715003794 Rev. B
155