Vial Presence; Syringe Assembly - Waters Alliance e2695 Operator's Manual

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Spills and condensation in the sample compartment are routed to a waste
connection below the front panel.
You can load a carousel into the carrier while an injection is underway. An
open-door sensor prevents the carrier from automatically advancing while you
load the carousel (see

Vial presence

A light-emitting diode assembly searches for the presence of a vial before the
needle moves to the Draw position. This searching prevents air from being
accidentally injected into the solvent stream when a programmed vial position
is accidentally left empty. See the Verify vial presence parameter in
configuration parameters" on page

Syringe assembly

The syringe assembly is located behind a front panel door for ease of access
(see the
figure "Syringe assembly:" on page
a standard 250-µL syringe; as an option, you can install either a 25-µL or
2500-µL syringe. See
change the syringe. Note that the separations module serial number is located
on the syringe mounting plate.
"Load carousels" on page
91.
"Replace the syringe" on page
Sample management system overview
111).
34). The separations module uses
205, for the procedure to
July 29, 2013, 715003794 Rev. B
"Set
33

Advertisement

Table of Contents
loading

Table of Contents