Waters Alliance e2695 Operator's Manual page 73

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

I/O signals: (Continued)
Signal
Chart out
Switch 1–switch 4
Run stopped
The following table describes the parameter voltages for the chart out signals.
Chart out signal conditions:
Parameter setting at 0
Signal
volts (minimum)
Flow rate
0.000 mL/min
System loop
–344.7 kPa (–3.45 bar, –50
pressure
psi)
Sample loop
–0.69 kPa (–0.007 bar, –0.1
pressure
psi)
Description
Outputs the analog signal (specified in the current
separation method, or default separation method) to a
device such as an integrator or strip chart recorder (see
"Set I/O parameter values" on page
one of the following signals as the chart out signal:
• Flow rate
• System pressure
• Sample loop pressure
• Primary head pressure
• Composition (%A, %B, %C, %D)
• Sample temperature
• Column temperature
• Degasser vacuum
The table, below, lists the output signal conditions for
each of these parameters.
See
"Event switches" on page
An output indicating that the separations module
ceased operating because of an error condition or
operator request (pressing the Stop Flow button, for
example).
162). You can select
74.
Parameter setting at 2.000
volts (maximum)
10.000 mL/min
37920 kPa (380 bar, 5500
3454.3 kPa (34.5 bar, 501 psi)
July 29, 2013, 715003794 Rev. B
Signal connections
1
psi)
73