Waters Alliance e2695 Operator's Manual page 232

Separation module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

8 Diagnostic Testing and Troubleshooting
Syringe position parameters:
Syringe position
Empty
Home
Full
Control the GPV valve
The following table describes the valve positions available in the GPV
drop-down list.
Gradient Proportioning Valve (GPV) position parameters:
GPV position
Off
Solvent A
Solvent B
Solvent C
Solvent D
Control valves V1–V4
To control a valve:
1.
In the Motors and Valves screen, select the valve whose position you
want to change.
2.
Press any numeric key to toggle the valve between the open and closed
position.
In the figure on
Tip:
indicated by the valve symbol.
3.
Press the Enter key to save your changes.
232
July 29, 2013, 715003794 Rev. B
Description
Syringe is in the highest position to purge the
syringe.
Syringe is in the home (default) position
(approximately 20% full).
Syringe is in the lowest permitted position (80%
full) to draw in sample.
Description
GPV off.
GPV selecting solvent from reservoir A.
GPV selecting solvent from reservoir B.
GPV selecting solvent from reservoir C.
GPV selecting solvent from reservoir D.
page
231, valve V1 is in the open position, as